ELECTRIC STERILMATIC STM-E TYPE 'C' W/TEMPERATURE RECORDER
Adj. legs
4"
.
6"
II
t
_--+-_r
""
L
__
J
Drain Catch
PIm
1
I'·
Approx.
4·6
Ou.ns
BvCustom
••
. _
See
Note 5
) '\
United Electric
Temperature Recorder
No.
6001·6531·
Style "c" Bulb
'Coiled) Temp. Range
160
To
28QO
F W/6Ft. Cap Line-Charts.
,.
Ink Pt. No. 95·2608
:r::..:.~...,_
;;
-':;:,
See
Note 2
II
.1
'I
locate
E
lecHlc
60)(
As Close
To
Sterahll!f
As Possible And Per Local
E
lectr
.cal
Code'>
REQUIRED CONNECTIONS
A
Drain 1/2 FPT
or
5/8
Copper (See Note
11
8
Steam Exhaust Connections 3/8 I.P.S.-'Note
21
C
Electrical Connections (See Note
3)
Nominal Amps Per Wire
3 Phase - 4 Wire -
1 Phase -
3 Wire
208V 240V
208V 240V
21
34
41
58
208
(197·2191 or
240V
(220-240'
0
115V .
6O·CY
. 1PH
·2.7
Watts
Units must
be grounded
NOTES:
1.
An
air break must
be
provided if a unit drain line
is
run.
2. Vent exhaust
tQ
atmosphere
(112"
nominal line
15'0"
Ig.
mu.
with min. of bends).
8
,
is
actual connection,
but
must exit casing
at
B slope exhaust away from unit
to
avoid
conden·
sate re-entering.
3. The timer & exhaust solenoid valve are
on
120V.
The heaters run
on
230V
or
208V. Thus if a
neu'
tral
to
the
power circuit if
not
available a separate
120V line must
be
run.
4. A 115V line may
be
run
separately
into
'0'
for
the
recording
thermometer,
or
the
line can
be
run from
'0'
to
'C'
by
connecting
to
proper leads.
5. The recorder must be located within 4 feet
of
the
sterilizer preferably
on
a solid wall
but
can be
mounted
on
the
side
of
the
sterilizer.
17-0099-3