SEPTEMBER 26, 2014 3 STM-E, STM-EL, STM-EX & STM-ELX STERILIZERS
DOOR ASSEMBLY & ADJUSTMENT
STERILMATIC DOOR ASSEMBLY
The Door of the Sterilmatic has been engineered to es-
tablish a positive method of sealing the steam pressure
within the sterilizing cylinder. As steam pressure builds
up within the cylinder, the door seal will tend to become
more positive.
However, the door should be adjusted to make a good
initial seal between the door gasket and the door opening
without the added assistance of internal cylinder steam
pressure with the simple action of securing the door han-
dle down in a locked position, the door gasket should be
sufciently compressed against the door opening, all the
way around to prevent any steam leakage from occurring.
Figure 1
DOOR ADJUSTMENT
The Door Adjustment is Located in the Fulcrum Casting
at the base of the door opening. This adjustment employs
the use of a screw and locknut in order to adjust the Steril-
matic Door to a tighter closed position (to prevent steam
from leaking by the door gasket as pressure builds up), it
is necessary to loosen the locknut and back off the screw
at least one-quarter of a turn and re-tighten the locknut.
Figure 2
THE DOOR GASKET
Keep the gasket clean. With normal closing and locking
of the door assembly, a steam-tight seal should be made
between the door gasket and the door opening. This seal
cannot be maintained if particles of foreign matter are al-
lowed to accumulate upon either of the contacting sur-
faces.
If there is leakage by the door gasket before a steam
build-up within the steam chamber and leakage does not
stop when the sterilizer reaches sterilizing temperature
and pressure than regard the door assembly as improp-
erly adjusted. A re-adjustment must then be made of the
seal adjustment door screw.
To change the door gasket, remove the entire door as-
sembly as a unit. Discard the old gasket, replace it with
a new one (no cement is required), and reinstall the door
assembly. Make an operational check for leakage and ad-
just the door, if necessary.