SECTION 1 INSTALLATION INSTRUCTIONS
COLD WATER CONDENSER
ITEM
PART
NO. DESCRIPTION
QTY.
1 95-2119 Steam condensing unit 1
2 95-2219 Thermostat Box Assy. 1
3 95-0086 Exhaust line 1
5 15-7057 Copper tubing 3/8 OD
22.25”
6 10-1775 Rd. Hd. Mach. Screw, 1/4-20 2
7 10-2500 Lockwasher, 1/4 2
8 10-2308 Hex Nut, 1/4-20 2
9 10-1812 Rd. Hd. Mach. Screw, 10-32 2
10 10-2505 Lockwasher, 10 2
11 10-2340 Hex Nut, 10-32 2
12 95-4009 Front Template (7” Lg) 1
13 95-4010 Back Template (11” Lg) 1
SERVICE CONNECTIONS REQUIRED
A
1/2” IPS Cold Water Connection
B
1” IPS Drain Connection (See Note 3)
C
115V Elec. Connection 7/8 Ø knockout (cond. unit)
D
Electrical Connection
NOTES:
Unit must be grounded and all wiring to comply with local codes.
Pipe to open drain. Do not make solid connection to sewer.
Condensing unit to be installed as to have a back to front pitch.
1.
2.
3.
AMPS PER WIRE *STERILIZER
Phase 3 Phase 1 Phase
208V (197-219)
or
240V (220-240)
Volts 208 240 208 240
Amps 26 30 45 52
Close chamber door.
Set exhaust selector to INSTRUMENTS AND
PACKS (fast exhaust) or LIQUIDS (slow exhaust).
Set timer to 15 minutes. Cycle will go to completion
automatically.
NOTE: Cycle timer will not start until sterilizing
temperature is obtained.
3.
4.
5.