SECTION 1 INSTALLATION INSTRUCTIONS
REQUIRED CONNECTIONS:
A Drain - 1/2” (13mm) FPT of 5/8” (16mm) OD
copper (see note 1)
B Steam Exhaust Connection - 3/8” (10mm) IPS
(see note 2)
C Electrical Connection - (*see table below)
D Power Supply
*AMP/PHASE
MODEL kW Hz
Single Phase
Three Phase
208V 240V 208V 240V
STM-E(L) 9 60 45A 52A - -
STM-E(L) 12 60 - - 26A 30A
UNIT MUST BE GROUNDED.
MAIN SUPPLY VOLTAGE FLUCTUATIONS ARE NOT TO
EXCEED ± 10% NOMINAL SUPPLY VOLTAGE
NOTES:
An air break must be provided if a unit drain line is run.
Vent exhaust to atmosphere. B1 is actual connection,
but must exit casing at B.
IMPORTANT: Exhaust line must be vented to the out-
side to eliminate the exhaust steam and the accompa-
nying noise from entering the room. Use 1/2” (13mm)
copper tubing or suitable alternate. The overall height
and length of the line should not rise more then 4’ (1.2
meters) above the unit and exceed 15’ (4.5 meters) with
a minimum of bends. The line should slope downward
after leaving the sterilizer in order to ensure condensate
drainage.
IMPORTANT: Failure to comply with this outline will af-
fect the sterilization process.
When an exhaust condenser is supplied; the following
services must be provided: 1/2” (13mm) IPS cold water:
1” (25mm) IPS waste: 115V electrical line.